To Use a Sensor Outside of the Setting Range
The scaling function can be employed to use a sensor outside of the setting
range.
1 On the Channel Settings screen, select the channel on which to use the sensor
outside of the setting range.
2 Move the cursor to [Sensitivity] to set the sensor sensitivity.
Enter a multiplier value that will produce a settable value (0.1 to 10)
when multi-
plied by the actual sensitivity of the sensor.
3 Move the cursor to [Scaling], and set it to [On].
4 Set the conversion ratio to the sensitivity multiplier value.
Setting Example 1:
If the specified sensor sensitivity is 23.4 pC/(m/s
2
):
Set 1/2.34 as the sensor sensitivity multiplier, and 10 (pc/m/s
2
)) as the sensor
sensitivity.
Scaling is set as follows to display measured values multiplied by 1/2.34.
Setting Example 2:
If the specified sensor sensitivity is 0.05 pC/(m/s
2
)
Set the sensor sensitivity multiplier to 2, and the sensor sensitivity to 0.1 (pc/m/
s
2
)).
Scaling is set as follows to display measured values multiplied by 2.
• Conversion ratio setting [Ratio]
Conversion ratio: 10/23.4 = 0.4274E+00
Offset: 0.0000E+00
Units: m/s
2
• Two-point setting [2-Point]
Input 1: 2.3400E+01 Scale 1: 1.0000E+01
Input 2: 0.0000E+00 Scale 2: 0.0000E+00
Units: m/s
2
• Conversion ratio setting [Ratio]
Conversion ratio: 0.1/0.05 = 2.0000E+00
Offset: 0.0000E+00
Units: m/s
2
• Two-point setting [2-Point]
Input 1: 0.0500E+00 Scale 1: 0.1000E+00
Input 2: 0.0000E+00 Scale 2: 0.0000E+00
Units: m/s
2