User Manual
4.1 Operation Overview
79
After turning power on, set the measurement criteria before measuring.
This chapter describes the settings related to logic inputs.
Refer to "7.3 Displaying Logic Waveforms" in the Instruction Manual for setting
details.
Measurement status and trigger settings are the same as for analog inputs.
Refer to the related chapters in the Instruction Manual.
Logic Input
Settings Chapter 4
4.1 Operation Overview
(Select the Function)
To record simultaneously with an analog signal, select a function that suits
the analog input.
When recording only logic signals, the function can be set freely.
Select the measurement status and
channels
(For Memory Function)
[Use Ch] page on Status Settings screen
Enable (set to On) the logic channels to be used.
See "4.2.1 Selecting Channels to Use" in the Instruction Manual
Set up input channels
[Logic] page on Channel Settings screen
Select whether to enable (On) or disable (Off) the input waveform for each
logic channel probe.
(Default setting: All Off)
Select waveform colors as desired.
See "7.3 Displaying Logic Waveforms" in the Instruction Manual
Set trigger criteria
[Logic] page on Trigger Settings screen
To apply a trigger event for measurements, set the logic trigger for each
channel.
See "6.8 Triggering by Logic Signals (Logic Trigger)" in "Chapter 6 Trigger
Settings" in the Instruction Manual
Waveform display settings
[Logic] page on Sheet Settings screen
Select the channels to appear on the waveform display.
(Default setting: All On)
Set the waveform display position and height for each channel as needed.
See "7.2 Setting the Screen Layout of the Waveform Screen (Sheet Set-
tings Screen)" and "7.3 Displaying Logic Waveforms" in the Instruction
Manual
Start Measurement