5.2 Analog Input Section
89
Temperature input
Measurement ranges 10, 20, 50, 100°C/div
Measurement input range K: -200 to 1350°C E: -200 to 800°C
J: -200 to 1100
°C T: -200 to 400°C
N: -200 to 1300°C R: 0 to 1700
°C
S: 0 to 1700
°C B: 300 to 1800°C
Temperature measurement
accuracy
K, E, J, T, N
R, S
B
±0.1% f.s.±1.0°C, ±0.1% f.s.±2.0°C (-200 to 0°C)
±0.1% f.s.±3°C
±0.1% f.s.±4°C (effective measurement range: 400 to 1800°C)
Reference junction compen-
sation
Selectable internal or external
Reference junction compen-
sation accuracy
±0.1% f.s.±1.5°C (with internal reference contact compensation and input terminal in
state of temperature equilibrium)
Temperature characteristic ±0.05%f.s./°C (sensor: K, E, J, T, N, with drift compensation mode OFF)
±0.25%f.s./
°C(sensor:R,S,B, with drift compensation mode OFF)
±0.04%f.s./
°C(all sensors, with drift compensation mode ON)
Frequency characteristic DC to 1 kHz dB
Data update rate: 250 μ70% (with drift compensation mode OFF)
1 s±20% (with drift compensation mode ON)
Input terminals 2-terminal terminal block
Input resistance 5.1 MΩ±5%
Low-pass filter OFF, 5±50%, 500±50% (Hz) -3 dB
+1