MEMORY HiCORDER 8860-50,8861-50 Recorders / Digital Oscilloscope New REC&MEM Function New Recording Logger and Oscilloscope The HIOKI MEMORY HiCORDER 8860 Series now sports vastly enhanced functionality. With new mouse and keyboard support, these models feature personal computer-like operability, accelerated by new internal high-speed hardware that provides simpler, faster operation.
As an Oscilloscope As a Data Logger Record Waveforms in Any Situation Model 8860-50/8861-50 Series Debut!
Capabilities and Features See page 4 20 MS/s high-speed sampling Up to 32 isolated input channels (high-speed signals) See page 5 Real-time saving to hard disk Dual-timebase sampling for simultaneous fast and slow monitoring Multi-channel logging on up to 128 channels (low-speed signals) See page 6 See page 6 Enhancements from Earlier Models 8860 and 8861 l A faster CPU greatly enhances instrument operability and response. l Three USB 2.
Reliably capture waveform anomalies buried within normal signals - Memory (Digital Oscilloscope) Function - Memory Function for High-Speed Waveform Monitoring A/D conversion Isolation Input signal Write on memory A/D conversion Display Thermal Printer Isolation Input signal Input signal Using the same operating principle as a digital oscilloscope, data is recorded to the expanded internal memory at high speed. Sampling rate is up to 20 MS/s (50-ns period) for all channels simultaneously.
Capture High-Speed Signals by Triggering During Slow Recording - New REC&MEM Function and Real-Time Saving Simultaneous Long-Term Monitoring and Transient Recording (REC&MEM) n Transient waveform recording that is impossible with a pen recorder The new REC&MEM function can record high-speed waveforms such as intermittent noise by applying a trigger while recording long-term fluctuations just like a pen recorder.
The Next Generation Hybrid Recorder - A single instrument provides both oscilloscope and data logger functions Installing a Scanner Module Creates a Multi-Channel Data Logger Recording slowly changing physical values such as temperature has been performed by plotting recorders and hybrid recorders (combined numerical value and analog graph recording), and is currently performed by data loggers. On the other hand, for high-speed waveform observation, only an oscilloscope (or MEMORY HiCORDER) can be used.
Accurately capture waveforms with diverse parameters - Advanced trigger function - Trigger during capturing and search after capturing Set multiple triggers on a single channel The trigger function allows you to set diverse parameters to detect a particular waveform anomaly during capturing. Setting the pretrigger mode allows you to monitor the pre-trigger waveform. This is useful for analyzing the cause of the anomaly.
Convert the time domain to the frequency domain for analysis - FFT analysis function - FFT analysis function The single-channel FFT function is used in spectrum analysis. The twochannel FFT function analyzes transfer functions. The octave analysis function is used in acoustic analysis. The signal source for FFT analysis is a section obtained from the waveforms captured in the MEM function (the required number of pieces of data for FFT analysis are 1000 to 20,000 points).
Specify an analysis point Waveform in the MEM function Perform FFT calculation on the waveform from the MEM function When performing FFT analysis on the data obtained by measurement with the MEM function, you can use the jog shuttle to specify an analysis point and view the calculation results on the same screen. Unlike with the conventional 8855 and 8841 models, you do not need to switch between the MEM function and FFT function screens to set the starting point of calculation.
Remote control with the Internet browser - LAN/USB, calculation function - n Remote control and automatic saving to a shared folder The 8860 series allows for remote control using the Internet browser on the computer. When you register access to a shared folder on a computer on the network, you can store and load data to and from HUB the shared folder on the 8860-50 or the 8861-50 file screen.
n Main unit Specifications Basic specifications 8860-50 (max. 4 input modules) 8861-50 (max. 8 input modules) Basic specifications 8860-50 (max. 4 input modules) Plug-in input modules Max. 32 analog channels (max. 128 channels with scanner unit) + 16 logic channels (standard configuration) Backup functions *3 Factory installation only - Input type/number of channels Plug-in input modules Max. 16 analog channels (max.
n Main unit Specifications Recorder functions Print/display section *6 Printer functions are available when optional printer unit is installed Display *6 Recording paper *6 Recording width *6 Paper feed density 10.4 inch TFT color LCD (SVGA, 800 × 600 dots) RECORDING PAPER 9231: 216 mm (8.50 in) × 30 m (98.43 ft), thermal paper roll (when using A4-size the printer unit 8995) RECORDING PAPER 9234: 112 mm (4.41 in) × 18 m (59.
n Main unit Specifications Additional features General Measurement parameter printing, cursor measurement, scaling, current clamp setting, comment input, screen hard copy, list/gauge, start condition hold, auto setup, auto save, remote control (start/stop/ print control), auto range, over-range indication, VIEW function, key lock, level monitor, vernier function, offset cancel, event marker input, waveform search function, report printing n Options specifications (sold separately) DC/RMS UNIT 8959 F
n Options specifications (sold separately) options common to Models 8861-50/8860-50/8861/8860/8842/8841/8835-01/8835/8826/8720 CONVERSION CABLE 9318 (to connect 9270 to 9272, 9277 to 9279 and 8940) CONVERSION CABLE 9319 (to connect 3273, 3273-50 and 8940) Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm, approx. 290 g (10.
n Options specifications (sold separately) Perform the same functions on the computer Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 30 cm (0.98 ft), approx. 150 g (5.3 oz) Note: The unit-side plug of the 9320-01 and 9327 is different from the 9320.
Configuration of options Note: Options described as ”factory-installed options” must be specified when ordering and cannot be installed by the user. Note: Product names appearing in this catalog are trademarks or registered trademarks of various companies. Input modules Input cables are not supplied. Please purchase the appropriate cable for the intended application. Logic signal measurement Install by inserting into the main unit. Can be replaced by user.