Data Sheet

15
LOGIC PROBE 9320-01/9327
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh; accuracy guaranteed
for 1 year)
Function
Detection of voltage signal or relay contact signal for High/Low state recording
Input
4 channels (common ground between unit and channels), digital/contact input,
switchable
(contact input can detect open-collector signals), input impedance:
1MΩ
(with digital input, 0 to +5V), 500kΩ or more (with digital input, +5 to
+50V)
, pull-up resistance: 2kΩ (contact input: internally pulled up to +5V)
Digital input threshold
1.4V/2.5V/4.0V
Contact input
detection resistance
1.5kΩ or higher (open) and 500Ω or lower (short), 3.5kΩ or higher (open)
and 1.5kΩ or lower
(short), 25kΩ or higher (open) and 8kΩ or lower (short)
Response speed
9320-01: 500ns or lower, 9327: detectable pulse width 100ns or higher
Max. allowable input
0 to +50V DC
(the maximum voltage that can be applied across input pins without damage)
n
Options specifications (sold separately)
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 30 cm
(0.98 ft), approx. 150 g (5.3 oz)
Note: The unit-side plug of the 9320-01 and 9327 is different from the 9320.
LOGIC PROBE 9321-01
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh; accuracy guaranteed
for 1 year)
Function
Detection of AC or DC relay drive signal for High/Low state recording
Can also be used for power line interruption detection
Input
4 channels (isolated between unit and channels), HIGH/LOW range switching
Input impedance: 100kΩ or higher
(HIGH range), 30kΩ or higher (LOW range)
Output (H)
detection
170 to 250V AC, ±DC (70 to 250V )
(HIGH range)
60 to 150V AC, ±DC (20 to 150V) (LOW range)
Output (L)
detection
0 to 30V AC, ±DC (0 to 43V)
(HIGH range)
0 to 10V AC, ±DC (0 to 15V) (LOW range)
Response time
Rising edge 1ms max., falling edge 3ms max.
(with HIGH range at
200V DC, LOW range at 100V DC)
Maximum allowable
input voltage
250Vrms (HIGH range), 150Vrms (LOW range) (the maximum voltage
that can be applied across input pins without damage)
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 1 m (3.28 ft),
approx. 320 g (11.3 oz)
Note: The unit-side plug of the 9321-01 is different from the 9321.
Cable length and mass: Main unit cable 1.3 m (4.27 ft), input section cable 46 cm
(1.51 ft), approx. 350 g (12.3 oz)
Perform the same functions on the computer
n
MEMORY HiVIEWER 9725
1) Application software enables you to perform
the same data analysis on a Windows
computer as on the MEMORY HiCORDERs
8860 series.
2) No confusion, because the screens
appearing on the computer are identical to
those of the 8860 series.
3) Functions identical to those of the 8860
series, such as waveform processing
calculation, run on the computer.
MEMORY HiVIEWER 9725
Compatible devices
MEMORY HiCORDER 8860-50, 8861-50, 8860, 8861
Supplied Media
One CD-R disc
Operating environment
Computer running under Windows 2000/XP, Vista 32-bit type
File loading
Readable data formats : Only for 8860 series data (.MEM, .REC,
.FFT, .SEQ, .IDX, .SET)
Maximum file size : 2 GW
File saving
Saved contents: measurement data (binary and ASCII), (partial
saving of the area between cursors A and B), setting conditions,
screen image (BMP, PNG), and calculation results
Display
Waveform display: 1-, 2-, 3-, 4-, 6-, and 8-split screen, horizontal,
vertical, consecutive scroll, and zoom in/out along the time
axis, move the zero position, zoom in/out, setting of variables
independently for each channel
X-Y-axis composite display
(for the MEM function only): 1-, 2-, and
4-split display, dot/line interpolation, composite area can be
specified
Numerical display: digital values of waveform data can be displayed
Display sheet: 16 sheets
Display channel count
(per sheet): 32 analog channels, 16 logic
channels, 16 calculated waveforms, 8 X-Y-axis composite waveforms
Cursor function: vertical cursor, horizontal cursor, trace cursor, two
cursors (cursor A and cursor B), time and voltage display
Clipboard copy: images on the waveform screen can be transferred
to the clipboard
Print
Supported printer: printer compatible with the OS
Print format: waveform image (1-, 2-, 3-, 4-, 6-, 8-, and
16-split), numerical print, report format, list print, calculation
results, screen image
Print area: the entire area, area between cursors A and B
Print preview: supported
DIFFERENTIAL PROBE 9322
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh, after 30 minutes of
warm-up time; accuracy guaranteed for 1 year)
Function
For high-voltage floating measurement, power line surge noise
detection, RMS rectified output measurement
DC mode
For waveform monitor output, frequency characteristics: DC to 10MHz
(±3dB), amplitude accuracy: ±1% of full scale (at max. 1000V DC), ±3%
of full scale
(at max. 2000V DC) (full scale: 2000V DC)
AC mode
For detection of power line surge noise, frequency characteristics: 1kHz to 10MHz ±3dB
RMS mode
DC/AC voltage RMS output detection, frequency characteristics: DC, 40Hz to
100kHz, response speed: 200ms or less
(400V AC), accuracy: ±1% of full scale
(DC, 40Hz to 1kHz), ±4% of full scale (1kHz to 100kHz) (full scale: 1000V AC)
Input
Input type: balanced differential input, input impedance/capacitance:
H-L 9MΩ/10pF, H/L-unit 4.5MΩ/20pF, Max. rated voltage to earth:
when using grabber clip 1500V AC/DC
(CAT II ), 600V AC/DC (CAT III ),
when using alligator clip: 1000V AC/DC
(CAT II), 600V AC/DC (C AT III )
Maximum allowable input
2000V DC, 1000V AC (CAT II), 600V AC/DC (CAT III)
Output
Voltage divider for 1/1000 of input, BNC connectors (output switchable
for 3 modes DC, AC, RMS)
Power source
(1) Connect the AC ADAPTER 9418-15, (2) Connect to the PROBE POWER
UNIT 9687 via the POWER CORD 9248, (3) Connect to HiCORDER logic
terminal via the POWER CORD 9324 and CONVERSION CABLE 9323,
(4) Connect to the F/V UNIT 8940 via the POWER CORD 9325.
DC POWER UNIT 9684
Note: Factory-installed option, built in on the bottom case of the main unit
Rated input voltage
12V DC (input range : 10 to 16V DC)
Power requirements
200VA (printer used)
Note: Only one of either the DC Power Unit 9684 or Probe Power Unit 9687 can be installed at any one time. Please contact your
HIOKI representative if concurrent installation is required.
PROBE POWER UNIT 9687
Note: Factory-installed option, built in on the bottom case of the main unit
No. of powerd channels
8 Channels, Output current: Up to 3 A total (Total combined
current consumption of connected probes should be no more than 3 A)
Compatible current probes
3273 (0.25 A max. current consumption): exclusive, up to 8 probes
3273
-50 (0.47 A max. current consumption): exclusive, up to 6 probes
3274 (0.46 A max. current consumption): exclusive, up to 6 probes
3275 (0.60 A max. current consumption): exclusive, up to 5 probes
3276 (0.44 A max. current consumption): exclusive, up to 6 probes
Compatible voltage probes
9322 (0.15 A max. current consumption): exclusive, up to 8 probes
Note: Only one of either the DC Power Unit 9684 or Probe Power Unit 9687 can be installed at any one time. Please contact your
HIOKI representative if concurrent installation is required.
Dimensions and mass: approx. 315.8 (12.43in) W × 29 (1.14in) H × 244.4 (9.62in) D mm,
approx. 1.25 kg (44.1oz) Accessories: None
Dimensions and mass: approx. 315.8 (12.43in) W × 18.2 (0.72in) H × 244.4 (9.62in) D mm,
approx. 570 g (20.1oz) Accessories: None
n
PC Software Specifications
Note: With use of the 8860-50/8861-50 , Wv ver 1.25 or later
Wave Viewer (Wv) Software (Application disk CD-R, bundled accessory)
Functions
• Simple display of waveform file
• Text conversion: convert binary data file to text format, with
selectable space or tab separators in addition to CSV, and
specifiable section, thinning available
• Display format settings: scroll functions, enlarge/reduce dis-
play, display channel settings
• Others: voltage value trace function, jump to cursor/trigger
position function
Compatible PC
operating systems
Windows 95/98/Me, Windows NT 4.0 (SP3 or later), 2000, XP
FlexPro (data analysis and presentation software)
Overview
Gather large data sets from multiple channels measured in different
timebases
Analyze and view results using the easy-to-navigate tabbed window
interface
Powerful search function lets you shift through large sets of
accumulated data of virtually any size
Functions
Over 100 analysis functions in both the time-domain and frequency-
domain, including:
Fourier Spectral Analysis
Time Frequency Spectral Analysis
Harmonic Analysis
Shock Response Analyses
Active Folder model for setting up a template of both diagrams and
functions that update every time a new signal is captured
Data
Presentation
Images in all common formats can be imported or exported
Export analyzed data either in traditional .CSV, .RTF or .TXT formats
or to 3rd party software to accommodate existing database structures
Easily arrange objects such as 2-D and 3-D diagrams, result tables and
company logos in a report or on website
Prepared report templates are automatically updated when new
measurements are made
Make cursor measurements, scrolling and zooming directly from the
report
Data Analysis and Presentation Software
n
Features
FlexPro is a powerful data analysis and presentation
software for importing and organizing data from the
8860-50/8861-50 Series.
Note: Product Company: Weisang GmbH (Germany)
Contact: Email: info@weisang.com
http://www.weisang.com/