4.1 Pre-Measurement Inspection
49
Before measuring, be sure to read the "Operating Precautions" (⇒ p. 5) and "Chapter 3
Measurement Preparations" (⇒ p. 37).
Basic Operations Chapter 4
4.1 Pre-Measurement Inspection
Do not use if damage is present, as
you could receive an electric shock.
Replace the damaged items.
Metal Exposed
Is the insulation of the probe or connection
cable to be used damaged, or is bare metal
exposed?
When using probes and connection cables
When using a clamp
Is the clamp cracked or damaged?
Yes
1
No Metal Exposed
2
To
No
2
Before using the instrument the first time, verify that it operates normally to ensure that the no
damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki
representative.
Peripheral Device Inspection
To
If damage is evident, request re-
pairs.
Yes
Is damage to the instrument or input mod-
ules evident?
Instrument and Input Module Inspection
When turning power on
Do the fans rotate and the Hioki logo
appear on the screen?
No
2
The power cord may be damaged
,
or the instrument may be damaged
internally. Request repairs.
Does the Initial or Waveform screen
appear?
No
Yes
The instrument may be damaged in-
ternally. Request repairs.
Nothing appears, or
the display is abnor-
mal
Yes
Inspection complete