Quick Start Guide

4.4 Actual Measurement and Analysis
56
2. Select the appropriate function
Function Menu
Opening Screen Case
Press the F1 [MEM] (Memory Function) key.
Using the CURSOR keys, move the cursor to the
Function menu, and press the F1 [MEM] key.
See "Function Selection" ( p. 29)
Waveform or Setting Screen Case
3. Set measurement configuration
2 Press the SUB MENU keys to select the [Status] menu item.
3 Press the SHEET/PAGE keys to
select the [Basic] page.
4 Set the timebase to [5 ms/div].
This is the time per division on the hor
i-
zontal axis.
Sampling rate = timebase (s/div) / 10
0
(samples per division)
5 Set the recording length to [50 div].
Set the recording time in number of div
i-
sions.
(Recording time = timebase (s/div) × r
e-
cording length (div))
Now use the CURSOR keys to mov
e
among the setting items, and press th
e
F1 to F8 keys to select the desired se
t-
ting.
Make other settings as occasion demands.
Supplement
When F8 [Page / ] appears, more selections are
available. Press the F8 key to display the
additional selections.
In the memory function case, sampling speed is 1/
100th the timebase. Each division contains 100
data samples. More detailed analyses are
possible when the timebase is set smaller.
Enable auto-ranging if you cannot determine the
appropriate range (
p. 51).
1 Press the SET key.
• When using multiple channels, verify that
appears with the channel to be used on the [Use
Ch] page.
See: "Setting the Timebase (Horizontal Axis)" ( p. A2) in
this manual
"4.2.1 Selecting Channels to Use",
"4.2.2 Setting the Timebase (Horizontal Axis) and
Sampling Rate",
"4.2.4 Setting the Recording Length (number of divi-
sions)"
in the Instruction Manual
See "Chapter 4 Measurement Configuration Settings" in the Instruction Manual