Quick Start Guide

4.4 Actual Measurement and Analysis
57
4. Set up the input channel
Check This!
Indicates the
type of module
(unit) selected.
3 Select the module [Unit1] and chan
-
nel number [Ch1] to be set.
4 Make appropriate input channel set
-
tings such as the measuremen
t
range.
Make other settings as occasion demands.
Setting Item Setting Value
(Verify) [Module]:Analog, Mode: Voltage
Range(/div)
20 V (per division)
(Each vertical division repre
-
sents 20 V)
Coupling
DC (default setting)
5 Make appropriate waveform displa
y
settings.
Make other settings as occasion demands.
See "7.1 Making Input Waveform Display Set
-
tings (Analog Waveforms)" in the Instruction
Manual
Setting Item Setting Value
Wave Disp
On (default setting)
Mag
(Magnification)
x1/2
Position 50% (default setting)
Supplement
Determining the range:
Measurement range is denoted as value per division
on the vertical axis.
Setting a measurement range causes the measure-
ment area and display area of the level monitor (on
the left side of the screen) to be correspondingly
changed.
It is convenient to watch the indicated display area
on the level monitor while setting the range.
6 Select the [Zero-Adjust] button.
Execute zero-adjust 30 minutes after turning power on.
Proper warm-up time depends on the particular input mod-
ule. Verify the specifications in the Input Module Guide.
A message appears when zero-adjust is fin-
ished.
1 Press the SUB MENU keys to select the [Channel] menu item.
2 Press the SHEET/PAGE keys
to select the [One Ch] page.
See "Chapter 5 Input Channel Settings" in the Instruction Manual and "Chapter 3 Input Channel Settings" in the
Input Module Guide
Now use the CURSOR keys to mov
e
among the setting items, and press th
e
F1 to F8 keys to select the desired se
t-
ting.
Level Monitor
To enter a comment for an individual channel:
A comment entered in the comment entry field ap-
pears on the waveform.
See "5.2 Adding Comments" in the Instruction Manual
To convert the displayed units for measurements
taken with a clamp or sensor, set the Scaling func-
tion.
See "5.4 Converting Input Values (Scaling Function)" on
the Instruction Manual