Quick Start Guide

5.5 Data Saving
72
Saving Waveforms
*2
See "10.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" in the Instruction Manual
1 Press the F2 [Waveform] key.
2 Select the data saving format.
3 Select the area to be saved.
4 Select the channels to save.
5 (when [Binary] saving format is selected)
Select to save by divisions depending on file siz
e.
See "Appendix 2.2 Waveform File Sizes" in the
Instruction Manual
6 Press the F1 [Execute] key.
Binary
To be able to reload data into the instru
-
ment (Default setting)
Text *
1
To be able to load data into a PC
Whole
Save all measurement data.
(Default setting)
A-B
Save data in the area demarcated by
the A/B cursors.
Displayed Ch
Saves the channels on all sheets fo
r
which waveform display is enabled
[On]. (Default setting)
All Ch
Saves all stored channels.
Off
Saves as one file without divisions.
(Divisions)
If the data is larger than the specified
recording length, it is divided and saved
in multiple files.
*1. When [Text] is selected as the save
type, data is saved with thinning.
Examples
2: Save every other data point.
(
xxx...)
5: Save one of every five data points.
(
xxxxxxxxxxxx..)
(
: saved, x: not saved)
"Shared Settings in Selection Save" ( p. 71)
Use the CURSOR keys to move the cursor to the
setting item, and select the desired setting contents
with the F1 to F8 key.
*2: [Target Blocks]
This is displayed when using the Memo-
ry Division function.
When [All Blocks] is selected, the index
file (SEQ) and waveform data of all used
blocks are saved as a batch.
About Memory Division:
"4.3.3 Dividing Memory" in the
Instruction Manual
Note:
Data remaining in memory is saved by
manual saving with the Real-time sav-
ing function.