The following two setting methods are available:
• Using the operating keys
• Using the TIME/DIV key (settable regardless of cursor position)
The data refresh rate is not allowed to exceed the maximum sampling rate of the
input module.
Example: Using an input module with maximum sampling rate of 1 MS/s (up to
1M samples per second). 1 MS/s = 1 µs/S (1 µs sampling period)
When the [Sampling Speed] is set to [50 ns/S], data is refreshed once each µs.
The maximum sampling rate of the input module being used can be verified on
the Config (Configuration) screen ("12.3.6 System Configuration List" (⇒ p.
357)). Also, when sampling at different rates, if the recording time determined by
the specified recording length is shorter than the Timebase 2 sampling rate, no
data is sampled on Timebase 2.
Timebase 1
Timebase 2
Recording Length (Recording Duration)
Sampling Period
Timebase Setting: Using the TIME/DIV Key
To open the screen: Press the SET key → Select with the SUB MENU keys →Status Settings screen
See Screen Layout (
⇒ p. 27), To set from the Waveform screen (⇒ p. 108)
Select slower
timebase
The value changes with each key press.
Select faster
timebase