4.2 Setting Measurement Configuration (Status Settings Screen)
91
4
Chapter 4 Measurement Configuration Settings
3
2
Normally, select [INT].
"Appendix 4.1 Sampling" (
⇒ p. A49)
To control sampling by an external signal
,
select [EXT] (⇒ p. 394).
Memory Function case:
Operating Key Procedure
1
SHEET/PAGE
Select the [Basic] page.
2
Select the sampling clock.
CURSOR
Move the cursor to the [Sampling Clock] item.
F1
Select [INT] (Internal). (default setting)
3
Select the timebase.
CURSOR
Move the cursor to the [Timebase] item.
F1 to F8
(Switch Display: F8)
Set the time per division (timebase) on the hori-
zontal axis.
The sampling rate changes accordingly. (you can
change it by the [Sampling Speed] setting)
5(default setting), 10, 20, 50, 100, 200, 500 μs/div
1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5 min/div
Timebase and Sampling Rate Settings: Using the
Operating Keys
To open the screen: Press the SET key → Select with the SUB MENU keys →Status Settings screen
See Screen Layout (
⇒ p. 27), To set from the Waveform screen (⇒ p. 108)
1
1
2
Recorder Function case
Operating Key Procedure
1
Select the timebase.
CURSOR
Move the cursor to the [Timebase] item.
F1 to F8
(Switch Display: F8)
Set the time per division (timebase) on the hori-
zontal axis.
2
Set the sampling rate.
CURSOR
Move the cursor to the [Sampling Speed] item.
F1 to F8
(Switch Display: F8)
Set the sampling rate.
The range of choices depends on the selected
timebase.
10(default setting), 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5, 10, 30 min/div, 1 h/div
(With Model 8958 installed: 50 ms/div to 1 h/div)
100 ns, 1 ms, 10 ms, 100 ms, 1 ms, 10 ms, 100 ms,
1s /S (Select a period that is 1/100 of the timebase
or less)
About sampling period:
"Appendix 4.4 Recorder Function Values
"
(⇒ p. A51)