4.2 Setting Measurement Configuration (Status Settings Screen)
92
Description Measuring with the Recorder Function
When the following timebase values are selected, displayed waveforms are
compressed in the horizontal (time axis) direction as shown.
50 ms/div
x1/2, 20 ms/div x1/5, 10 ms/div x1/10
When the recording length [Shot] is to set [Cont] (Continuous), the timebase
must be set to at least 20 ms/div. Faster timebase settings are not available.
When the Model 8958 16-Ch Scanner Unit is installed, the timebase can be
set between 50 ms/div and 1 h/div.
If the sampling rate is set too fast, when the input waveform amplitude is
small, the difference between maximum and minimum values may become
quite large as a result of sudden impulses such as noise. To prevent such
phenomena, select a slower sampling rate or enable the input module’s low-
pass filter ( p. 111).
See"Appendix 4.4 Recorder Function Values" ( p. A51)
This applies to the Memory function only.
Different sampling rates can be set for Timebase 1 and Timebase 2. The follow-
ing channels can be set to Timebase 2.
Channels on which you want to measure with a slower sampling rate than that
of Timebase 1 ( p. 93).
Channels on the Model 8958 16-Ch Scanner Unit when used together with
another input module (Timebase 2 is then set exclusively for the 8958) (
p. 94).
4.2.3 Setting Different Sampling Rates
Setting a slow sampling rate for Timebase 2 results in longer preparation time
prior to the start of storage.