User Manual
4.2 Setting Measurement Configuration (Status Settings Screen)
95
4
Chapter 4 Measurement Configuration Settings
Set the length (number of divisions) to record each time data is acquired.
The following methods and settings are available:
• Fixed recording length [Fixed]: select from the fixed recording lengths
(⇒ p. 95).
• Set arbitrary recording length [User]: set an arbitrary recording length in
units of divisions (⇒ p. 97).
• Continuous [Cont]: records continuously (Recorder Function only) (⇒ p. 98).
Recording Length and Data Samples
• Memory Function
Each division of the recording length consists of 100 data samples. The total
number of data samples for a specified recording length = set recording length
(divisions) × 100 + 1.
• Recorder Function
Each recording length division = 100 pairs of data points, with each pair com-
posed of two samples: the maximum and minimum measured values within
each sampling period.
See "Appendix 4.4 Recorder Function Values" (⇒ p. A51)
4.2.4 Setting the Recording Length (number of divi-
sions)
To change recording length while measuring
Recording length can be changed on the Waveform or Settings screens. The re-
cording length becomes effective at the time the setting is changed.
See Modifying the Waveform screen view: "4.4 Setting Measurement Configuration on
the Waveform Screen" (
⇒ p. 108)
3
2
Setting a Fixed Recording Length (Fixed Shot)
To open the screen: Press the SET key → Select with the SUB MENU keys →Status Settings screen
See Screen Layout (
⇒ p. 27), To set from the Waveform screen (⇒ p. 108)
Operating Key Procedure
1
(with the Memory function)
SHEET/PAGE
Select the [Basic] page.
2
Select the setting method for recording length.
CURSOR
Move the cursor to the [Shot] item.
F1
Select [Fixed].
3
Set the recording length.
CURSOR
Move the cursor to the [Fixed Shot] (Fixed re-
cording length) item.
F1 to F8
(Switch Display: F8)
Select the length of waveform to be acquired (re-
cording length).
Displayed recording time and maximum
recording length are linked to the record
-
ing length setting.
1