4.2 Setting Measurement Configuration (Status Settings Screen)
98
See "Appendix 2.4 Memory Capacity and Maximum Recording Length" ( p. A37)
"Appendix 2.3 Timebase and Maximum Recordable Time" (
p. A32)
Setting Continuous Recording (Cont)
To open the screen: Press the SET key Select with the SUB MENU keys Status Settings screen
See Screen Layout (
p. 27), To set from the Waveform screen ( p. 108)
Operating Key Procedure
1
CURSOR
Move the cursor to the [Shot] item.
2
F3
Select [Cont] (Continuous).
Indicates the maximum number of divi-
sions remaining in internal memory tha
t
can be retraced after recording has
stopped.
Real-time printing
Real-time printing is not available when the timebase is 20 to 200 ms/div, even
if Auto Print (real-time printing ( p. 301)) is [On]. Of course printing can still
be performed manually after finished measuring ( p. 303).
Up to 5,000 divisions of data can be internally recorded by the instrument
(with the Model 9715 Memory Board installed).
Data is not saved internally during measurement. Data remaining in memory
is saved when measurement is manually aborted.
When using the Model 9684 DC Powr Unit, or when using the Model 8995-01
A6 Printer Unit to print numerical values, real-time printing is not available at
timebase settings of 500 ms/div or 1 s/div.
Measuring beyond the maximum recording length
When [Cont] recording is selected and measurement continues beyond the
recording length, the remaining recording time displayed on the Waveform
screen becomes negative after the end of the recording time (zero). (except
when display of both date and time is enabled) ( p. 336)
Timebase setting with [Cont] recording
The timebase can be set to any value from 20 ms/div when the recording length
is set to [Cont]. If the timebase has been set to 10 ms/div, selecting [Cont]
recording length forces it to 20 ms/div.