User Manual

4.3 Acquiring Waveforms Using the Utility Functions
105
4
Chapter 4 Measurement Configuration Settings
2
3
1
Memory Division: Display Settings
To open the screen: Press the SET key Select with the SUB MENU keys Mem Div Settings scree
n
See Screen Layout ( p. 35), To set from the Waveform screen ( p. 213)
Operating Key Procedure
1
To display any block on the Waveform screen
Set the display blocks
(This can also be set on the Waveform screen.( p. 213))
CURSOR
Move the cursor to the [Display Block].
F1 to F8
Set the number of blocks to display on the Wave-
form screen after measurement.
2
To display multiple blocks as overlaid waveforms
Enable the Reference Block function
CURSOR
Move the cursor to the [Ref Block].
F2
Select [On].
(When Reference Blocks are enabled [On])
Select whether to reference every block
CURSOR
Move the cursor to the row number of the Refer-
ence Block to select its block number.
F7 or F8 Enables (On) or disabled (Off) Reference Blocks.
When enabled, the frame of the selected block is
red.
F1 to F8
To overlay all waveforms, select the [All On] but-
ton.
3
To display every block as its waveform is acquired
Enable the Trace Waveform display
CURSOR
Move the cursor to the [Wave Display].
F2
Select [On].
Viewing Memory Division waveforms on the Waveform screen
See "8.12 Viewing Waveforms in Every Display Block (Memory Division)"
(
p. 213)
Off Reference Blocks are not displayed
(default setting)
On Reference Blocks overlay Display blocks
on the display.
All Off Disables all block references.
All On Enables all block references.
Off The waveform of the specified Display
Block is displayed after recording the
specified number of Used Blocks.
(default setting)
On Waveforms are displayed one block at a
time as they are acquired at each trigge
r
event.
Display Block (Green)
Reference Block (Red)
Measurement data is recorded at the
colored positions.
When the Display Block is 5 and the
Reference Block is 3
Reference Block No.
Enabling the Trace Waveform display
lengthens dead time.
About Dead Time:
See: "Difference Between Dead Times
During Normal and Memory Division
Recording" (
p. 106)
When Using Auto Save
When disabled, displayed images are no
t
saved.
Even if the Roll Mode is enabled (othe
r
than Off), it is not usable when the Trace
Waveform display is disabled.
Reference Block Selection
Reference Blocks can also be selected
and deselected in the [Reference Block
]
item on the [List] display.
See: "Getting Details on Each Block" (
p. 106)