User Manual

4.3 Acquiring Waveforms Using the Utility Functions
106
Description Difference Between Dead Times During Normal and Memory Division
Recording
Getting Details on Each Block
The trigger time and measurement status of each block can be viewed on the [List]
screen.
Select F2 [List].
Block No.
A
block can be selected by the CURSOR keys or the F5 to F8 keys. Yo
u
c
an move the cursor to the Reference Block column to set a block’s on/off sta
te
a
s a Reference Block.
To switch block waveforms on the Waveform screen
You can use the SHEET/PAGE keys to select blocks to be displayed.
In the default state, the SHEET/PAGE keys switch Display Sheets. You can
change the function of the SHEET/PAGE keys to [Block Switching] on the System
- Environment Setting Screen.
See "12.2.5 Specifying SHEET/PAGE Key Operations" ( p. 340)
When both printer recording (Auto Print) and Auto Save are set for continuo
us
triggering [Cont]
Dead Times
Times during which sampling is inhibited due to internal
processing, printing or saving
Recording Length
Anomalous phenomena occurring during
dead times are not detected.
W
hen the Trace Waveform Display is disabled (Off) during Memory Division re
-
c
ording
12 34 56
Dead Times
The waveform data of each recording length is recorded in one block.
When recording with Memory Division, dead time is shorter than with normal recording.
When Trace Waveform Display is enabled, dead time is longer.
Recording Length