4.3 Acquiring Waveforms Using the Utility Functions
107
4
Chapter 4 Measurement Configuration Settings
The times during which sampling is inhibited (dead time) due to display and
recording processing after each block of data has been acquired are about 8
ms.
When measuring a parameter other than voltage or current with the Model
8940 F/V Unit, dead time is about 230 ms.
When using the Model 8958 16-Ch Scanner Unit or Timebase 2 sampling,
dead time may be longer, depending on the Timebase 2 sampling speed set-
ting.
When the Trace Waveform display is disabled, even if the Roll Mode is
enabled (other than Off), the Roll Mode function is unusable.
When triggering occurs very often, pressing the STOP key may not stop mea-
surement until enough data has been acquired to fill the blocks specified for
use.