User Manual

5.4 Converting Input Values (Scaling Function)
118
When saving text or numerical calculation
results
In certain cases, entered characters may
be changed when data is saved.
( p. 282)
Operating Key Procedure
1
SHEET/PAGE
Select the [One Ch] page.
(Setting can also be done on the [Scaling] page.)
(
p. 125)
2
Enable the Scaling function.
CURSOR
Move the cursor to the [Scaling] item.
F2
Select [On] (Default setting: Off).
3
Set the display format for numerical values.
CURSOR
Move the cursor to the [Format] item.
F1 to F8
Select either choice.
4
Specify the physical units.
CURSOR
Move the cursor to the [Units] item.
F1 to F8
Enter the physical unit name.
(Up to 7 characters)
See "Entering Text and Comments" (
p. 65)
5
Select the scaling conversion method.
CURSOR
Move the cursor [Ratio] or [2-Point] (item current-
ly selected).
F1 to F8
Select either choice.
6
Enter the numerical values for conversion.
When you have selected [Ratio] (set conversion ratio
and offset)
CURSOR
Move the cursor to each of the [Ratio] and [Off-
set] items.
F1 to F8
Enter numerical values in each field.
See "Entering Numbers" (
p. 64)
When you have selected [2-Point] (set input values for
two points and the values after conversion)
CURSOR
Move the cursor to the [Input 1], [Scale 1], [Input
2] and [Scale 2] items.
F1 to F8
Enter numerical values in each field.
See "Entering Numbers" (
p. 64)
When entered, converted values are displayed in the specified phys-
ical units on the level meter. You can check that the setting values
are correct (
p. 119).
Decimal Displays decimal values.
Exponential Displays exponents in multiples of 3.
Ratio Specify by conversion ratio.
2-Point Specify by two points.
-9.9999E+9 to 9.9999E+9
-9.9999E+29 to 9.9999E+29
2
3
4
5
6
Setting Scaling
To open the screen: Press the SET key Select with the SUB MENU keys Channel Settings scree
n
See Screen Layout ( p. 30)
When setting by two points
The point values can be set to the current
input values displayed on the monitor.
Example:
Decimal format .............1.2345 mV
Exponential format ....... 1.2345E-03 V
"Scaling Methods" ( p. 117)
You can check that the conversion setting
values are correct. (
p. 119)
If you want to enter the current input value
as it is to Inputs 1 and 2, select F7 [Monitor
Value].
[Ratio] Setting
[2-Point] Setting
When using a clamp, scaling can be
performed automatically. ( p. 120)