5.5 Verifying and Setting All Channels from a List
125
5
Chapter 5 Input Channel Settings
Move the cursor to the [Preset] button, and select F1 [Preset].
The [Zero-Position] dialog appears.
To set the zero position of all channels at once: execute Preset
Reset the zero positions of all channels to the default value
.
Set the zero positions of all channels to 50%.
Set the zero positions of all channels to 0%.
To make more detailed settings:
1. Set the reference position.
2. Set the number of steps from the set reference position a
s
a percentage (%).
3. The zero position is set to the specified step.
Details of zero position:
"7.1.2 Setting the Waveform Display Position (Zero Posi
-
tion)" ( p. 166)
The setting changes when you select any button.
Scaling Settings List: [Scaling] Page
To open the screen: Press the SET key Select with the SUB MENU keys Select the [Scaling] pag
e
with the SHEET/PAGE keys
Using the CURSOR keys, move the cursor to the [Ch] column.
To set Scaling:
Select F1 [All Settings] and set from the dialog. (Each setting can be made
when the cursor is moved to the setting item.) Setting choices are the same
as on the [One Ch] page.
See"5.4 Converting Input Values (Scaling Function)" ( p. 117)
To copy settings from one channel to another:
Select F2 [Copy].
This mark indicates settings
are valid.
Select the Scaling method.