User Manual
6.2 Setting Workflow
131
6
Chapter 6 Trigger Settings
Trigger settings can be made on the Trigger Settings or Waveform screen.
Settings choices for each item are function-dependent.
6.2 Setting Workflow
Trigger Mode Settings
(
⇒ p. 132) Set whether to continue to accept triggers after measuring.
(function-dependent)
[Single], [Repeat] or [Auto]
Pre-Trigger Settings
(
⇒ p. 134) Set the amount to record before and after a trigger point.
• [% Setting] or [DIV Setting]
• Trigger priority
Trigger Timing Settings
(
⇒ p. 138) Set the start and stop timing for recording by triggers.
[Start], [Stop] or [Start & Stop]
Trigger Type Setting
Make trigger source-related settings.
• Analog trigger (
⇒ p. 140)
(Level, In-Window, Out-of-Window, Period, Glitch, Slope or
Voltage Sag trigger)
• Logic trigger (⇒ p. 153)
• Timer trigger (
⇒ p. 156)
• Manual trigger (⇒ p. 159)
• External trigger (⇒ p. 160)
Setting AND/OR Between
Trigger Sources
(
⇒ p. 133) Select AND/OR logical combination of analog, logic, exter-
nal and timer triggers to be applied.
[AND] or [OR]
Start Measurement
Press the START key to start measuring.
The green LED lights.
(However, operation differs according to the setting for START
key behavior (
⇒ p. 337).)
Data acquisition starts when trigger criteria are met.
To stop measurement: press the STOP key.
Press once: recording stops at the end of the specified record-
ing length.
Press twice: recording stops immediately.