6.6 Setting Trigger Timing
139
6
Chapter 6 Trigger Settings
Example: When the trigger type is Level Trigger, Level = 0.000 V, and Slope
= (rising)
Trigger timing
[Start]
[Stop] [Start & Stop]
Records for specified
recording length
Press START key to record
Records until a trigger
occurs
TT
START key
Recording starts when
a
Start trigger event occurs
Records until a Stop trigge
r
occurs
TT
T
he above sequences repeat when the trigger mode is
[
Repeat].