User Manual

6.7 Triggering by Analog Signals
142
Setting by Dialog ([Analog Trigger] dialog)
123
4
5
6
7
Move the cursor to each item, and make the setting.
After making settings, select the [Close] button to accept the changes.
F1
Move the cursor to a trigger to set in the [No
.]
column, and select F1 [All Settings] to ope
n
the dialog.
Setting Items Setting Description
1
No.
Trigger No.
2
Type
Select the trigger type ( p. 140).
3
Ch
(only with the [Expanded] setting)
Select the module (Unit) and Channel No. to which this trigger
applies. (1-1 = Unit 1, Channel 1)
4
Level
Set the signal level (threshold voltage) for triggering.
A trigger is applied when the signal crosses this level.
5
Slope
Select the slope (input signal rising or falling ) for triggering.
6
Filter
Set the filter width (trigger filter) for triggering.
Prevents noise from causing false triggers (
p. 143).
7
Parameters
(Event, Timing)
Make other settings.
Specify the event count (only with the [Expanded] setting) for
triggering (
p. 143).
When [Start & Stop] is selected for trigger timing (
p. 138) with
the Recorder function, select which triggers to use to start and
stop recording.