6.8 Triggering by Logic Signals (Logic Trigger)
155
6
Chapter 6 Trigger Settings
Operating Key Procedure
5
Select the trigger detection method.
CURSOR
Move the cursor to the [Detect] item.
F1 to F8
Select either choice.
6
When using the Memory function, or when using the
Recorder function with [Timing] set to [Start & Stop]
Set the triggers to start or stop recording.
CURSOR
Move the cursor to the [Timing] item.
F1 to F8
Select either choice.
Level Triggering occurs when the criteria ar
e
met. If the criteria are already met whe
n
measurement starts, the trigger is ap
-
plied.(default setting)(See Note)
Edge Triggering occurs when the specified cr
i-
teria are met (after not being met). If th
e
criteria are already met when measure
-
ment starts, no trigger is applied until afte
r
the criteria cease to be met and are the
n
met again.
Start Set the trigger to start recording.
(default setting)
Stop
Set the trigger to stop recording.
"About trigger timing" (⇒ p. 138)
T
Lch A 1 1
Lch A 2 0
Lch A 3 X
Lch A 4 0|1
[Level] (Trigger source: when [AND])
Start Measurement
T
Lch A 1 1
Lch A 2 0
Lch A 3 X
Lch A 4 0|1
[Edge] (Trigger source: when [AND])
Start Measurement
Setting external and timer triggers with the [AND] trigger source setting
In the following cases, triggering occurs in the same way as with the [Edge] set-
ting even when [Level] trigger detection is selected.
If logic trigger criteria have been met before an external or timer trigger is
applied, no triggering occurs. When external and timer trigger criteria have been
applied, and after they have been subsequently released, triggering occurs once
all trigger criteria are met again.
(Example: when logic and timer triggers have been set)
T
rigger Detection [Detect]: Level
T
rigger Source [Trigger]: AND
Lch A 1 1
Lch A 2 0
Lch A 3 X
Lch A 4 X
Timer Trigger
T
Start Measurement
T
Start Times
Interval