6.12 Making Trigger Settings on the Waveform Screen
161
6
Chapter 6 Trigger Settings
The following trigger criteria settings can be made on the Waveform screen.
Press the SUB MENU keys to select available setting items.
Trigger Mode
Pre-Trigger (Memory function and FFT function only)
Analog Trigger (settings depend on the trigger type)
Use the CURSOR keys to move the cursor to each setting item, and select your
choice with the F keys.
6.12 Making Trigger Settings on the Waveform
Screen
1
Memory Function
4
Analog Trigger Settings
57
10
689
1
Memory Function
23
"4.4 Setting Measurement Configuration on the Waveform Screen" ( p. 108)
Measurement Configura-
t
ion and Trigger Settings
SUB MENU key
11
1457
10
Recorder Function
89
11
6
Setting Items Description
1
Function
2
Trigger Mode ( p. 132) Sets the trigger mode.
Single, Repeat or
Auto
Single or Repeat
Single, Repeat or
Auto
3
Pre-Trigger ( p. 134) Sets pre-triggering.
-100 to 100%
(In steps of 1%,
or divisions)
(None)
-100 to 100%
(In steps of 1%,
or divisions)
4
Analog Trigger No. Selects the trigger number.
5
Analog Trigger Type
(
p. 140)
Selects the analog trigger type.
6
Unit and Channel No. Selectable only when [Expanded] is selected.
7
Trigger Level Set the signal level (threshold voltage) for triggering.
8
Trigger Slope Select the slope (input signal rising , falling or both rising and falling ↑↓) for triggering.
9
Trigger Filter ( p. 143) Sets the filter width (trigger filter) for triggering.
10
Events ( p. 143) Sets the event count for triggering. (only with the [Expanded] setting)
11
Timing ( p. 138)
Set the timing for triggered recording.
For Recorder function, set this when trigger timing is set to [Start &
Stop].
________