User Manual
7.2 Setting the Screen Layout of the Waveform Screen (Sheet Settings Screen)
174
The default setting assigns channels in the order of input module installation.
However, with the Memory function or Real-time saving function, only those
channels enabled for use [On] can be assigned.
See "4.2.1 Selecting Channels to Use" (⇒ p. 85)
7.2.6 Assigning Display Channels to Graphs (Analog
Channels)
Operating Key Procedure
1
Select the channels to display on the Sheet.
SHEET/PAGE
Select the [Analog] page.
CURSOR
Move the highlight cursor to the No. to set, and
then to the [Ch] column.
F1 to F8
Select the channel number of the unit (module) to
display on the Sheet.
To not display the channel, select F3 [Off].
2
Select the display graph (when Split-Screen is set to
[2 Graphs] or more).
CURSOR
Move the cursor to the [Graph] column.
F1 to F8
Select the graph number in which to display.
Verify the pattern display for the graph number.
1
2
Indicates the input type for selected chan
-
nels.
Analog Channel Assignment
To open the screen: Press the SET key → Select with the SUB MENU keys →Sheet Settings screen
See Screen Layout (
⇒ p. 168)
Setting from a dialog
Move the cursor to the
[No.] column of the channel to be set, and select F1 [All
Settings]. A dialog appears. Set each item, then select the [Close] button.
If “Storage Off” appears
A selected channel is disabled ([Off]) on the [Use Ch] page of the Status Settings
screen. To display, set the channel to [On] and measure again.
If “Display Off” appears
Waveform display setting on the Channel Settings screen is disabled ([Off]). To
display the waveform on the Waveform screen, set it to [On].