User Manual

7.4 Composite Waveforms (X-Y Waveforms)
182
Making Partial
Composites
Make a partial composite after specifying the composite range within normal
waveforms using the A/B cursors.
See "8.7 Specifying a Waveform Range" ( p. 193), "8.8 Cursor Values" ( p. 195)
Method 1 View the normal waveform display and the partial composite waveform on
a separate sheet
Set the Sheet Settings screen as follows:
Sheet 1 (S1)
Display Type: [Waveform]
Sheet 2 (S2)
Display Type: [X-Y Comp] or [Wave & X-Y]
Composite Area: [A-B]
Also make the required settings for the X-Y composite such as composite
channel selections.
Display Sheet 1 (S1) on the Waveform screen, and specify the waveform range
for the composite using the [Vertical] or [Trace] mode of the A/B cursors. The
sheet displayed on the Waveform screen can be switched by the SHEET/PAGE
keys.
The composite range can be changed from Sheet 1.
Making X-Y composite settings from a dialog
Move the cursor to the [No.] column to be set, and select F1 [All Settings].
A dialog appears. Move the cursor to each item and select with the F keys.
To reset graph settings
Move the cursor to the [Reset Graphs] button, and select F1 [Reset Graphs].
Graph numbers are reassigned sequentially from the top of the setting column.
F1
Horizontal cursors cannot be used to specify the range for partial composites.
When one cursor is used, the X-Y composite range is from the cursor to the
end of the waveform.
Sheet 1 (S1)
Normal Waveform
Sheet 2 (S2)
Partial Composite Waveform