User Manual
8.10 Fine Adjustment of Input Values (Vernier Function)
211
8
Chapter 8 Waveform Screen Monitoring and Analysis
Fine adjustment of input voltage can be performed arbitrarily on the Waveform
screen. When recording physical values such as noise, temperature and accel-
eration using sensors, amplitude can be adjusted to facilitate calibration.
8.10 Fine Adjustment of Input Values
(Vernier Function)
Normal Display Vernier Function Enabled
1.2 V input is displayed as 1.0 V
1.2 V
1.0 V
Press the DISP key to display the Waveform screen.
Using the CURSOR keys, move th
e
cursor to an item in the dialog, an
d
press the F keys to select it.
1
4
Select the channel to
adjust.
Adjust while viewing the waveform.
Adjustment Range: between 50 and 200% of origina
l
waveform (magnification/compression ratio not dis-
played)
: Magnify waveform
: Compress waveform
Press the RANGE/POSN knob to
display the [CH SET] dialog.
2
3
• The Vernier function cannot be applied to the calculation waveforms.
• Vernier adjustments cannot be verified on printed waveforms or lists.