User Manual
8.14 Searching a Waveform
219
8
Chapter 8 Waveform Screen Monitoring and Analysis
Operating Key Procedure
Setting Logic Channel Search Criteria
10
Select the criteria number to use.
CURSOR
F1 to F8
Move the cursor to the [No.] item.
Select a number for this set of search criteria,
from Analog 1 to Analog 8.
11
Select the timebase to search (if measured using two
timebases).
CURSOR
F1 to F8
Move the cursor to the [Axis] item.
Select either choice.
12
Select the trigger search criteria for logic channels
Set the trigger search criteria for each channel.
To combine different search criteria with logical [AND] or [OR],
repeat steps 10 to 12 to specify all necessary criteria.
Search
13
CURSOR
F1
Move the cursor to the [Exec] button.
Starts searching.
[Searching] appears on the Status bar, and the
locations where search criteria are satisfied are
displayed.
Search markers ( indicate locations where
search criteria are met.
To search for more instances
F2
Select [Next].
To move a cursor to a found location
F4
or F5 Select [Move Cur A] or [Move Cur B].
14
Close the search function.
F8
Select [Close].
11
10
12
Logic channel
search criteria
Trigger filterÅiFilter WidthÅj
Trigger pattern
13 14
Timebase 1 Searches a Timebase 1 channel (de
-
fault setting).
Timebase 2 Searches a Timebase 2 channel.
S
Trigger search criteria settings are made
using the same procedures as the pre-
measurement trigger criteria settings.
See "6.8 Triggering by Logic Signals
(Logic Trigger)" (
⇒ p. 153)
Select ( ) [Auto Close] to
have the dialog close auto-
matically after searching.