User Manual

10.3 Saving Data
265
10
Chapter 10 Saving/Loading Data & Managing Files
Settings such as measurement configurations can be saved to storage media by
pressing the SAVE key.
In addition, multiple instrument setting states ("settings data") can be stored in
internal instrument memory and reloaded.
Settings data can be loaded automatically at power-on (Auto Setup function) (
p. 281).
10.3.6 Saving Settings Data
2
Saving Settings Data: Saving to Storage Media
To open the screen: Press the SET key Select with the SUB MENU keys Save Settings screen
See Screen Layout (
p. 38)
Operating Key Procedure
1
Set manual save ( p. 263).
Set the save destination.
2
Select the save type.
CURSOR
Move the cursor to the [Save Type] item.
F1
Select [Settings].
Press the SAVE key after setting the measurement configura-
tion to save.
The settings data that was set is saved to the specified storage media upon
pressing the key.
See "10.6 Examples of Saving Data: Reading Data on a PC" (
p. 282)
[SAVE Key] Page
The data of all settings configured in the
Settings screen can be saved.
(However, the communication settings
cannot be read. If the communication set
-
tings are required, save or print the display
screen. )
See
"10.3.10 Optionally Selecting Display
Screens & Saving (SAVE Key)" (
p. 274
)
"11.4 Making Manual Print (PRINT Key
Output) Settings" (
p. 303)
When you want to load the settings data automatically at power-on (Auto
Setup function):
If you create a STARTUP.SET file for auto setup, the settings data can be loaded
automatically from the storage media at power-on.
See "10.5 Saving & Loading Auto Settings File (Auto Setup Function)" ( p. 280)
In addition, previously saved settings data can be reloaded when the instrument
is turned on.
See "Saving Settings Data: Internal Saving" ( p. 266)
"Select the data to load: Loading from the instrument" (
p. 278)