User Manual

10.4 Loading Data
275
10
Chapter 10 Saving/Loading Data & Managing Files
Waveform data or settings data saved to storage media can be loaded on the
instrument.
Furthermore, if you create a STARTUP.SET file for auto setup ( p. 281), the
settings data can be loaded automatically from the storage media at power-on.
See "Creating a Settings File for Auto Setup" ( p. 281)
"Automatically Loading Settings Data (Auto Setup)" ( p. 281)
Multiple setting states can be stored in the instrument and later reloaded. In the
same way, settings can also be automatically reloaded when power is turned on.
See "Select the data to load: Loading from the instrument" ( p. 278)
Data Not Loadable on the Instrument
Data saved in text, BMP, or PNG format.
Data saved on devices other than the 8860 and 8861.
When the data is loaded, the file name appears on the bottom left of the wave-
form screen. The file name is displayed until the START key is pressed.
File Types
See "10.2 Data Capable of Being Saved & Loaded" ( p. 252)
To load waveform data in a batch ( p. 279)
An index file is necessary to read files as a batch.
Load any of the following types of index files.
The index file is required to load files saved as a batch.
10.4 Loading Data
File Extension
IDX
Loads all saved files that were divided at a specified recording length.
SEQ
(Memory function only)
Loads all saved files that were saved as individual blocks by Memory Divi-
sion.
RSI
(Real-Time Saving Function only)
Loads files saved using the Real-Time Saving function.
When the data is loaded from the storage media:
The storage media needs to be inserted before it can be selected.
If the data on the storage media is to be modified, make sure write protection is
disabled before you insert the storage media.
If the “Cannot load this file” message appears:
Check the format of the selected file. The instrument can load waveforms and
settings data saved in binary format.