User Manual
10.4 Loading Data
278
Previously saved settings data can be loaded from storage media (File screen)
or from internal memory (System Settings Status screen).
Loadable settings data file: “file name”.SET
10.4.2 Loading Settings Data
Select the data to load: Loading from the storage media 
To open the screen: Press the FILE key → File screen
Operating Key Procedure
1
Select the data to load (⇒ p. 276).
2
F1
(Page1/3)
Select  [Load].
A confirmation dialog box appears.
3
F1
Select [Execute].
The selected settings data is loaded on the in-
strument.
To cancel loading
Select 
F2 [Cancel].
If F1 [Load] is not displayed, press the F8
key to switch to [Page 1/3]. 
To load settings data automatically:
If you create a STARTUP.SET file for auto setup, the settings data can be loaded
automatically from the storage media at power-on.
See "10.5 Saving & Loading Auto Settings File (Auto Setup Function)" (⇒ p. 280)
Select the data to load: Loading from the instrument
To open the screen:  Press the DISP key→ Press the F7 [System] key→ Select with the SUB MENU key
s 
→ Setting Configuration screen
Operating Key Procedure
1
CURSOR
Move the cursor to the Setting No. to load.
2
F1
Select [Load].
A confirmation dialog box appears.
3
F1
Select [Execute].
The selected settings data is loaded in the instru-
ment.
To cancel loading
Select F2 [Cancel].
Setting Number 
A mark beside the No. indicates that the
setting state is stored.
Loading stored settings data automatically
A setting state can be automatically loaded when turning power on.
Enable [Auto Setup] (set to On), and set the
Setting No. to the number of the settings data to