User Manual

11.6 Setting the Print Content
317
11
Chapter 11 Printing
Time Axis Magnification and Compression Settings
To open the screen: Press the SET key Select with the SUB MENU keys Print Settings screen
See Screen Layout (
p. 39)
Operating Key Procedure
1
SHEET/PAGE
Select the [Print Items] page.
2
CURSOR
Move the cursor to the [Mag/Comp] item.
F1 to F8
Select the display type.
3
When [No Screen Link] is selected
Set the magnification or compression ratio.
CURSOR
Move the cursor to the magnification or compres-
sion ratio field.
F1 to F8
Set the magnification or compression ratio for the
timebase.
No Screen
Link
Print using the magnification or com
-
pression ratio set here.
Screen Link Print using the magnification or com
-
pression ratio set for the waveform
screen (default setting).
(Memory Function or Sampled waveform data
from the Real-Time Saving function is enabled)
x 10 to x 1/50000
(Recorder Function or Whole waveform data from
the Real-Time Saving function is enabled)
x 1 to x 1/20000
Regardless of the magnification or com-
pression ratio set for the waveform
screen, the magnification or compression
ratio set here is printed.
The Recorder Function display magnification ratios x 4 and x 2 are valid only
for screen display. When printing waveforms, the magnification ratio x 1 gives
1 pixel per data point, which is the same resolution as the screen when
viewed at x 4.
During external sampling, depending on the timebase magnification ratio, the
number of data points per division may be a number with a decimal fraction.
When the waveform is printed, the decimal fraction is discarded, so that the
number of data points per division is an integral number.