12.3 Making System Settings
352
Operating Key Procedure
1
CURSOR
Move the cursor to the [Key Check] button.
F1
Select [Execute].
Operation keys appear.
2
All operation
keys
Press each operation key once or more.
The corresponding key is painted over.
Jog: Rotate to the left and right, one time or
more in each direction.
Shuttle: Rotate all the way in left and right
directions.
A/B knobs and RANGE/POSN knobs: Rotate
each knob to the left and right one or more
times. Press the inner side knobs.
The START key also functions as an LED light
check key.
The check is finished when you have operated all
the keys.
To cancel the check
Press the START and STOP keys simultaneously. The original screen reappears.
If you are using a mouse, you can click the right button and select [Exit].
Key Check
To open the screen: Press the DISP keyPress the F7 [System] keySelect with the SUB MENU key
s
Init Settings screen
See Screen Layout (
p. 48)
The key check does not finish if there is a problem that prevents even one of the
keys from being recognized. If this occurs, press the START and STOP keys
simultaneously to display the original screen.
There may be a malfunction in the instrument, so request repairs. If there is a
problem with the STOP or START key, you cannot return to the original screen.