12.3 Making System Settings
356
Zero-position adjustment of the 8958 16-Ch Scanner Unit with this instrument is
necessary in the following cases. Adjust one hour after powering on.
When the 8958 16-Ch Scanner Unit is installed in this instrument
(A message appears when this instrument boots, prompting you to perform
the adjustment.)
When the zero position has gotten out of alignment due to the passage of time
or changes in the environment
12.3.5 Adjusting the 8958 16-Ch Scanner Unit
Before executing the adjustment
Before adjusting the scanner unit, the + and - terminals of channels 1 and 9
must be shorted. For details, refer to "Scanner Unit Zero Position Adjust-
ment" in "2.2.7 Connecting to the Model 8958 16-Ch Scanner Unit" in the
Input Module Guide.
Short the + and - terminals of channels 1 and 9
CH1 CH9
+ +
Scanner Unit Adjustment
To open the screen: Press the DISP keyPress the F7 [System] keySelect with the SUB MENU key
s
Init Settings screen
See Screen Layout (
p. 48)
Operating Key Procedure
1
CURSOR
Move the cursor to the [Adjust Scanner Unit] but-
ton.
F1
Select [Execute].
The [Adjustment] dialog box appears.
2
F1 to F8
Select the scanner unit to adjust.
Check to be sure that the + and - terminals of
channels 1 and 9 are shorted on the selected
scanner unit.
F7
Select [Execute].
To cancel adjustment
Select
F8 [Cancel].
A processing message appears when you exe-
cute the adjustment.
The adjustment is complete when "Completed
normally" appears.