14.1 Connecting External Control Terminals
389
14
Chapter 14 External Control
Connecting the SYNC.OUT Jack and the EXT.SMPL Jack (BNC connector)
Connect the BNC plug of the connection cord to
the BNC jack on the instrument side.
Align the slots on the BNC plug with the guide pins on
the instrument-side jack, then push and twist the plug
clockwise until it locks.
Disconnecting BNC connectors
Twist the BNC plug counterclockwise and pull it out.
Lock
Instrument-side
guide pins
BNC plug slots
Connecting External I/O Terminals (Connector Blocks)
Cables to connect
Single strand
Multi-strand
10 mm
Connection procedure
1
2
3
• Recommended cables:
single strand diameter 1.0 mm (AWG18),
multi-strand 0.75 mm
2
(AWG20)
• Usable cables:
Single strand diameter 0.4 to 1.0 mm (AWG26 to 18),
Multi-strand 0.3 to 0.75 mm
2
(AWG22 to 20)
Strand diameter 0.18 mm or greater (per wire)
• Standard insulation stripping length:10 mm
• Button operation specified tool: Flat-blade screwdriver
(tip width 2.6 mm)
1 Push in the button on the connector with a flat-
blade screwdriver or other tool.
2 With the button held in, insert the cable into the
cable connection hole.
3 Release the button.
The cable is locked.