User Manual
14.2 External I/O
399
14
Chapter 14 External Control
Signals can be output when the results of evaluation of numerical calculations
are GO (pass) or NG (fail).
The default setting (factory default) is [Num Calc].
Evaluation result output period (for numerical calculations: 100 ms or greater
For one-shot measurements, the signal is saved. It returns to HIGH on the next
start.
For continuous measurements, the signal is saved until the next trigger event
occurs.
Example of a circuit for a warning device
14.2.5 GO/ NG Evaluation Output (GO/EXT OUT1)/ (NG/EXT OUT2)
Evaluation result is
GO or NG
Signal is output.
GO/EXT OUT1
NG/EXT OUT2
The GO/EXT OUT1 and NG/EXT OUT2 terminals can be used as GO/NG evalu-
ation output terminals (GO, NG) or as external output terminals for specific con-
ditions (EXT OUT1, EXT OUT2) (⇒ p. 401). They cannot be used for both
functions at the same time.
GO/ NG Evaluation Output Signals
Output signal Open collector output (with voltage output), active LOW
Output voltage range HIGH level: 4.0 to 5.0 V, LOW level: 0 to 0.5 V
Maximum input voltage -20 to +30 V 500 mA max 200 mW max
HIGH
4 to 5.0 V
LOW
0 to 0.5 V
GND
4.7 kΩ
4.7 kΩ
10 kΩ
5 V
GO, NG
Evaluation result output period
GO, NG
GND
5 V 5 to 24 V
Warning output
Relay
Reset
1 μF
HC74
1 to 5 kΩ
2SC1815 etc.
HC04
10 kΩ
D
Q
PR
CK
R