User Manual

14.2 External I/O
402
2 In the Ext Term (external terminal) screen, make settings for the [GO/EXT
OUT1] and [NG/EXT OUT2] external output terminals.
See For more information about other menu items
"14.2.5 GO/ NG Evaluation Output (GO/EXT OUT1)/ (NG/EXT OUT2)" (
p. 399)
The signal for the specified state is output.
Select the conditions under which the instru
-
ment outputs a signal.
Error Output a LOW level signal when an error occurs.
Busy Output a LOW level signal when the instrument cannot start an opera-
tion because it is starting another operation, saving data, printing, and
so on
Start Output a LOW level signal while instrument is starting.
Trigger Output a LOW level signal while instrument is waiting for a trigger, and
when a trigger event occurs.
Default setting: GO/EXT.OUT1: Numerical evaluation
NG/EXT.OUT2: Numerical evaluation
(To open the screen: Press the DISP keyPress the F7 [System] key
Select with the SUB MENU keys Ext Term Settings Screen)