User Manual

15.3 Memory Function
415
15
Chapter 15 Specifications
15.3 Memory Function
Timebase 5, 10, 20, 50, 100, 200, 500 μs/div
1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5 min/div
External sampling (100 S/div) allows simultaneous control of multiple instruments
Time axis resolution 100 points/div
Sampling period 1/100th of timebase, or external sampling
The timebase can be set according to sampling period
Two different sampling periods can be set
Recording Length Fixed or adjustable
The setting range depends on the capacity of installed memory and the number of chan-
nels enabled for use
("Appendix 2.4 Memory Capacity and Maximum Recording
Length" (
p. A37))
Fixed settings
25, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000, 50000, 100000, 200000,
500000, 1000000, 2000000, 5000000, 10000000
Adjustable settings
1 to 10240000 (in 1-div steps)
Screen and Printing
Settings
1, 2, 3, 4, 6, 8 or 16 screens (printer, excluding Model 8995-01 A6 Printer Unit)
Can be displayed sequentially, or split into selected widths (with some restrictions)
X-Y screens (1 or 4 screens) (X-Y and time axis screens can be combined)
Sheet display (up to 32 channels displayed per sheet)
Interpolation function Line (exc. X-Y), dot or line (with X-Y)
Recording line distinction 32 colors (four printing types)
Overlay function Provided
Maximum Recording Length [Divisions]
Installed Memory
(Words)
No. of Channels Used
8860
168421
8861
32 16 8 4 2
32M
64M 20,000 20,000 50,000 100,000 200,000
128M
256M 50,000 100,000 200,000 500,000 1,000,000
512M
1G 200,000 500,000 1,000,000 2,000,000 5,000,000
1G
2G 500,000 1,000,000 2,000,000 5,000,000 10,000,000
Maximum Recording Length [Divisions]
Installed Memory
(Words)
No. of Channels Used
8860
168421
8861
32 16 8 4 2
32M
64M 20,000 40,000 80,000 160,000 320,000
128M
256M 80,000 160,000 320,000 640,000 1,280,000
512M
1G 320,000 640,000 1,280,000 2,560,000 5,120,000
1G
2G 640,000 1,280,000 2,560,000 5,120,000 10,240,000