User Manual

15.3 Memory Function
416
Waveform compression
and magnification
Time axis
× 10, × 4, × 2, × 1
× 1/2, × 1/5, × 1/10, × 1/20, × 1/50, × 1/100, × 1/200, × 1/500,
× 1/1000, × 1/2000, × 1/5000, × 1/10000, × 1/20000, × 1/50000,
× 1/100000, × 1/200000, × 1/500000
Voltage axis
× 100, × 50, × 20, × 10, × 5, × 2, × 1, × 1/2, × 1/5, × 1/10
Waveform scrolling Left-right scrolling by Jog and Shuttle knobs
Waveforms can be viewed and scrolled before measurement finishes
(Roll Mode: restricts time axis and waveform compression)
Parts of the waveform already recorded can be scrolled into view while measuring
Auto Print On or Off: automatically prints recorded waveforms
(Selectable for whole waveform, or for cursor-defined selection)
Manual Print Available
The whole waveform or cursor-defined selection can be selected for printing by PRINT
key settings
Printout magnification can be set independently from display magnification
Selection printing Prints the waveform between A/B cursors (by PRINT key setting)
Smoothed printing Setting print quality to [Fine (Slow)] doubles print density in the time axis direction, pro-
viding smooth waveform printing (Only using the Model 8995 A4 Printer Unit)
Report Print Available
Login function Prints and displays measurement data as numeric values
Variable display function Provided (voltage axis)
Upper and lower limits and range position can be set
Variable settings can be linked to changes in voltage range settings
Zoom function Provided (split-screen display of whole waveform and magnified section is available)
X-Y Composites X-Axis: 8 channels, Y-Axis: 8 channels (8 composites)