User Manual
15.4 Recorder Function
417
15
Chapter 15 Specifications
15.4 Recorder Function
Time axis 10*
1
, 20*
1
, 50*
1
, 100*
1
, 200*
1
, 500 ms/div
1, 2, 5, 10, 30, 50, 100 s/div
1, 2, 5, 10, 30 min/div
1 h/div
*1. Real-time paper recording is not available with the faster ranges (10 ms to 200 ms/
div), although waveforms can still be recorded in memory and monitored on-screen.
Up to 5000 divisions (with Model 9715 Memory Board installed) of waveforms can
be stored before measurement stops. Also, if the Recording Length is set to other
than [Cont], simultaneous printing is available, so waveforms can be printed under
the following conditions:
With [Cont] Recording Length, 20 ms/div to 1 h/div
With Model 8958 16-Ch Scanner Unit installed, 50 ms/div to 1 h/div
Time axis resolution 100 points/div
Sampling period 100 ns, 1 μs, 10 μs, 100 μs, 1 ms, 10 ms, 100 ms, 1s
(not more than 1/100th of the selected timebase)
Recording length Fixed, User (Adjustable) or Cont (Continuous)
(setting range depends on the capacity of installed memory)
• Fixed settings
25, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000, 50000, 100000
• Adjustable settings
1 to 160000 (in 1-div steps)
Continuous setting is not available for 10 ms to 200 ms/div timebase settings when
printing
Timebase settings of 10 ms/div to 1 s/div are not available when printing numerical val-
ues on the Model 8995-01 A6 Printer Unit
Screen and printing
settings
1, 2, 3, 4, 6, 8 or 16 screens (printer, except on the Model 8995-01 A6 Printer Unit), can
be displayed sequentially, or split into specified widths (with some restrictions)
Sheet display (up to 32 channels displayed per sheet)
Interpolation function Line
Recording line distinction 32 colors (four printing types)
Waveform magnification
and compression
•Time axis
× 4*, × 2*, × 1, × 1/2, × 1/5, × 1/10, × 1/20, × 1/50, × 1/100, × 1/200, × 1/500,
× 1/1000, × 1/2000, × 1/5000, × 1/10000, × 1/20000
(* Screen display only. Printing is can be x 1 or more)
• Voltage axis
× 100, × 50, × 20, × 10, × 5, × 2, × 1, × 1/2, × 1/5, × 1/10
Maximum Recording Length [Divisions]
Installed Memory
(Words)
Other than the Model 8958 16-Ch
Scanner Unit
Model 8958 16-Ch Scanner Unit
8860 8861
Fixed
Adjustable or
Continuous
Fixed
Adjustable or
Continuous
32M 64M
5,000 5,000 1,000 1,000
128M 256M 20,000 20,000 5,000 5,000
512M 1G 50,000 80,000 20,000 20,000
1G 2G 100,000 160,000 20,000 40,000