Timebase • Measurement waveform
100, 200, 500 μs/div
1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30 s/div
1, 2, 5 min/div
•
Whole waveform (with auto setting function)
10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 100 s/div
1, 2, 5, 10, 30 min/div
1 h/div
May be limited by saving destination and number of channels
Time axis resolution 100 points/div
Sampling period Measurement waveform: 1/100th of the timebase
Whole waveform: same as measurement waveform
Save destinations MO drive, hard drive (HDD), LAN, PC Card or OFF (none)
Recording Length Maximum recording length: determined by available space at the save destination, the
file system, number of channels and whole waveform timebase
Length is set in units of divisions, up to the maximum recording length
Screen and Printing
Settings
1, 2, 3, 4, 6, 8 or 16 screens (printer)
Can be displayed sequentially, or split into selected widths (with some restrictions)
Sheet display (up to 32 channels displayed per sheet)
Recording line distinction 32 colors (four printing types)
LCD (display) While measuring: whole waveform
After measuring: selectable from whole waveform, measurement waveform, or both
whole and measurement waveforms displayed simultaneously (split-screen).
Printer Output When not measuring, the whole or measurement waveform can be printed as displayed
on the LCD
Zoom function Provided (when only a measurement waveform is displayed)
Report Print Available
Variable display function Provided (voltage axis, timebase)
Upper and lower limits and range position can be set
Variable settings can be linked to changes in voltage range settings
Login function Prints and displays measurement data as numeric values