15.7 Functions
422
Comment Entry Title comment
Comments for each channel
Comments can be printed at the zero-position of each channel at the left side of printed
waveforms
Comment printing for each channel using callouts on waveforms
Screen Image Capture
function
Provided (for printing and saving as BMP files)
List On or Off Prints setting conditions following waveforms
Gauge On or Off Prints before waveforms
Available for on-screen display
Grid Off, Normal, Fine, Normal (Dark), Fine (Dark), Time Axis, or T-Axis (Dark) (printout only)
Retain Start Condition
function
Provided
Retains continuity of timer trigger criteria
Auto Setup function Automatically loads settings from external storage media when turning power on
Auto Save function Provided
Remote control Control terminals to Start, Stop, Print and Save
Settings are provided to change operations
(2.5 V threshold, Active Low or Shorted Terminals)
Auto-Ranging Function Provided (Memory function only)
Automatically select the optimum timebase and voltage axis range
Error Display Displays the cause when an error occurs
Key-lock Keys (other than KEY LOCK) can be temporarily disabled
LCD Backlight On, Off (Auto-Off function)
Screen Saver On, Off (Auto function)
PRINT Key setting Provided
Print contents can be selected by pressing the PRINT key
(Screen linkage, whole waveform, between A/B cursors, pre- and post-trigger wave-
form, report, list, calculation results, screen image)
With the “Screen Link” setting, waveforms on the Waveform screen are printed by
pressing the PRINT key: either whole waveforms, or if the A/B cursors are enabled, just
the waveforms between cursors are printed
Lists (of settings) can be printed from screens other than the Waveform screen
SAVE Key settings Provided
Settings are provided to select storage media, save format, file name and saving area
by pressing the SAVE key
Level Monitor function Provided (Level bar, measurement values)
Monitoring is available while measuring and awaiting triggers
Logic display On, Off and Comments are available for each waveform
Any of 16 display positions can be selected for each block of four channels (L-Chs A to D)
Wide, Normal or Narrow logic waveform width (height) can be specified
Vernier function Fine adjustment of input voltage can be made arbitrarily (from 50 to 200% of original
input level)
Offset Cancel function Executing Offset Cancel causes the measured input value to be recognized as zero
Waveform search
functions
Search criteria can be specified as trigger criteria, specified time or peak value
TIME/DIV direct setting
function
The timebase can be changed using the special TIME/DIV key
Range and Position direct
setting function
The range and position settings of input modules can be adjusted using the special-pur-
pose knobs