User Manual

Appendix 1 Error Messages
A3
Appendix
34
Invalid key pressed (Overlay)
The key operation is prohibited because
Overlay is enabled (On).
"4.3.2 Overlaying Wave-
forms" (
p. 101)
36
No trigger has been set. Set trigger criteria.
"Chapter 6 Trigger Set-
tings" (
p. 129)
37
Invalid operation.
The operation is not available while pro-
cessing. Try again after processing finish-
es.
38
Invalid operation (measuring).
The operation is not available while mea-
suring. Try again after measuring finishes.
39
Invalid operation (printing).
The operation is not available while print-
ing. Try again after printing finishes.
41
Recording length is set to Continuous.
When the Recording Length (Shot) is set
to Continuous (Cont), real-time printing is
not available with fast timebase settings
"11.2 Print Methods and
Print Items" (
p. 299)
42
There is no calculation result.
There is no calculation result. Print results
after performing calculation.
Analysis Supplement
45
Out of range.
Check the valid setting range, and reset.
50
Roll Mode is not available.
The Roll Mode cannot be used when
Overlay is enabled.
"4.3.1 Displaying Wave-
forms During Recording
(Roll Mode)" (
p. 99)
54
Can not use (Averaging, Overlay, Wave
calculation).
Averaging, Overlay and Waveform calcu-
lation functions are prohibited when the
Roll Mode is set to [On] or [Auto].
"4.3.1 Displaying Wave-
forms During Recording
(Roll Mode)" (
p. 99)
56
Real-time printing is not available.
Recording Length is set to [Cont]. Real-
time printing is not available when the time
base of the Recorder function is 10 to 200
ms/div.
"Setting Continuous Re-
cording (Cont)"(
p. 98)
59
Can not use (Averaging, Wave calcula-
tion).
When Memory Division is enabled, Aver-
aging and Waveform Calculation func-
tions are not available. If enabled, these
functions are turned off.
60
No waveform data. Acquire waveform data.
61
Can not use (Roll Mode, Averaging,
Memory Division).
When Waveform Calculation is enabled,
Roll Mode, Averaging and Memory Divi-
sion functions are not available. If en-
abled, these functions are turned off.
64
Up to eight clamps can be used.
Up to eight channels can be used simulta-
neously when clamps are connected to
the Model 8940 with Model 9318 or 9319
Conversion Cables. Do not exceed this
clamp limitation. When using the Model
9325 connected to the 9322, up to eight
channels can be used for clamps and the
Model 9322 together.
Input Module Guide:
"3.5.5 Current Measure-
ment"
67
Too many measurement channels.
You have tried to use more channels than
the number enabled for use. Either in-
crease the number of channels enabled
for use, or turn unused channels Off.
"4.2.1 Selecting Chan-
nels to Use" (
p. 85)
Displayed Warnings
Msg
No.
Message Remedial Action Reference