Appendix 1 Error Messages
A5
Appendix
81
Perform initialization to start synchro-
nized measurements.
A setting was changed after starting syn-
chronous operation. Resynchronize by
pressing the [Synchronous Start] button
on the System screen of the master instru-
ment.
"14.2.4 Synchronized
Sampling Output
(SYNC.OUT)" (
p.
396)
82
Maximum sampling rate is limited to 1
MS/s.
The sampling rate is limited when using
synchronized sampling.
83
No channel selected for use. Select the channel(s) to use.
"4.2.1 Selecting Channels
to Use" (
p. 85)
84
Measurement aborted due to save pro-
cessing delay.
With the Real-Time Saving function, mea-
surement is aborted if the recording speed
at the save destination is too slow. Select
a slower timebase or reduce the number
of channels used.
85
Invalid search condition. Check the search criteria settings.
"8.14 Searching a Wave-
form" (
p. 215)
86
Recording length is too long. Check the recording length.
"4.2.4 Setting the Record-
ing Length (number of divi-
sions)" (
p. 95)
87
Cannot load measurement waveform.
Load an index file (.RSI) created by the
Real-Time Saving function. If the RSI file
cannot be loaded, a measurement wave-
form file (.RSM) may be damaged or miss-
ing.
94
No response from server.
Verify the network settings on the PC at
the connection destination.
501
An unexpected error occurred when ac-
cessing (file name).
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
502
(File name) was not found.
Verify that the file saving destination or file
to load is correct.
503
(File name) is an invalid path.
Verify that the file saving destination or file
to load is correct.
"10.2 Data Capable of
Being Saved & Loaded"
(
p. 252)
504
Too many open files. Cannot open (file
name).
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
505
Access to (file name) refused.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
506
(File name) has an invalid file handle.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
507
The current directory (file name) cannot
be deleted.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
508
Not enough free space in the directory
to create (file name).
Either delete files in the saving destination
directory, or change to another saving
destination.