Seek for (file name) failed.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
510
A hardware I/O error occurred while ac-
cessing (file name).
The storage media may be corrupted. Re-
place with new storage media.
511
A sharing violation occurred while ac-
cessing (file name).
Verify the settings (user name and pass-
word) of the shared destination.
512
A locking violation occurred while ac-
cessing (file name).
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
513
Not enough disk space for (file name).
Saving is not possible because of insuffi-
cient space on the storage media. Delete
files or replace the storage media. If mea-
suring, stop measurement, then replace
the storage media.
"10.7.3 Deleting Files &
Folders" (
p. 291)
514
Attempted to access past the end of (file
name).
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
520
(File name) may be corrupted.
The file may be corrupted. This file cannot
be used.
Displayed Warnings
Msg
No.
Message Remedial Action Reference
Displayed Errors
Msg
No.
Message Remedial Action
160
Measurement aborted (Cause)
(Cause) Real-time measurement and saving operations were
aborted due to an error. Remove the cause of the error.
170
Unsupported FPGA version detected.
Upgrade to a compatible FPGA.
Upgrade to an FPGA version that is supported by the application
program.
171
Unsupported Kernel version detected.
Upgrade to a compatible Kernel.
Upgrade to a kernel version that is supported by the application
program.
180
USB host controller malfunction de-
tected.
Have the instrument repaired.
181
Keyboard controller malfunction de-
tected.
Have the instrument repaired.
182
I/O FPGA malfunction detected.
Have the instrument repaired.
183
Storage FPGA malfunction detected.
Have the instrument repaired.
184
Bus bridge malfunction detected.
Have the instrument repaired.
190
Module power supply malfunction de-
tected.
Power to the modules was momentarily interrupted. If this oc-
curred while measuring, data may have been corrupted.
191
Clamp power supply malfunction de-
tected.
Power to the clamps was momentarily interrupted. If this occurred
while measuring, data may have been corrupted.