Appendix 1 Error Messages
A7
Appendix
Other Display Messages
These messages provide only supplemental information.
193
Module power supply malfunction de-
tected. Turn power off immediately.
Power to the modules is abnormal. Turn the instrument off imme-
diately, and have it repaired.
194
Clamp power supply malfunction de-
tected. Turn power off immediately.
Power to the clamps is abnormal. Turn the instrument off immedi-
ately, and have it repaired.
195
Fan malfunction detected. Turn power
off immediately.
A cooling fan has stopped. Turn the instrument off immediately,
and have it repaired.
Displayed Errors
Msg
No.
Message Remedial Action
Msg
No.
Message Remedial Action Reference
102
Verify Sheet settings.
Verify settings such as the channels to be
displayed on sheets. Verify the settings on
the Sheet Settings screen.
"7.2 Setting the Screen
Layout of the Waveform
Screen (Sheet Settings
Screen)" (
p. 168)
119
Unit (module) configuration has
changed. Verify each setting.
When an input module has been added or
replaced, verify settings on the Settings
screens (Status, Channel, Trigger and
Sheet). Pay particular attention to the dis-
played channels setting on the Sheet Set-
tings screen.
Input Module Guide:
"2.1 Installing Input Mod-
ules (Adding or Replac-
ing)"