3.1 Measurement Workflow
51
7
3
Chapter 3 Operation Overview
Refer to "Appendix 2.1 List of Default Settings" (⇒ p. A8) for default settings.
The default setting for Auto Save and Auto Print is Off (disabled).
Set the items indicated by white text within the boxes as needed. To simultaneously record logic waveforms,
also read "3.1.2 Logic Waveform Recording" (⇒ p. 55).
Operation
Overview Chapter 3
3.1 Measurement Workflow
3.1.1 Analog Waveform Recording
Procedure (asterisks (*) indicate settings
that can be changed while measuring) Overview and references
Installation & Connection
Install the input modules and cables required
for measurement.
See "Chapter 3 Measurement Preparations" in the Quick
Start Manual
"Chapter 2 Connections" in the Input Module Guide
Zero-Adjustment is required if the 8958 16-Ch Scanner Unit is
installed (⇒ p. 356).
See "2.2.7 Connecting to the Model 8958 16-Ch Scanner
Unit" in the Input Module Guide
Power On
Function Selection
Select the appropriate function.
See "Choosing the Appropriate Function" (⇒ p. 81)
• Memory Function
Records relatively fast signals from µs to minutes
• Recorder Function
Record relatively slow signals at low speeds from ms to
hours
• Real-Time Saving Function
Long-term measurement data can be saved to storage
media in real time (
⇒ p. 225).
• FFT Function
Using frequency analysis, spectral analysis and transfer
functions can be performed. Analysis is applied to data
measured with the Memory function.
(Refer to the Analysis Supplement for details of the FFT
function)
Instantaneous
Waveform Recording
Long-Term Recording
Memory
(Sampling point
recording)
Recorder
(Envelope recording)