4.2 Setting Measurement Configuration (Status Settings Screen)
87
4
Chapter 4 Measurement Configuration Settings
When using logic channels
The default setting is [On], but if insufficient space is available for the specified
number of channels to be used, some channels are set [Off]. In this case, set
unneeded channels [Off] or increase the set number of channels to use, and
then set the needed logic channels [On].
Decreasing the number of channels to be used below the number of chan-
nels set [On]
Channels are automatically set to [Off], starting with the lowest channel.
Using the Model 8946 4-Ch Analog Unit and logic channels
Maximum recording length is limited in the following conditions.
* Model 8860: 32 MWords, Model 8861: 64 MWords memory installed
No. of Chs to Use
Used Channels
Max. Rec.
Length*
Model 8946 4-Ch Analog Units
Logic channels
8860 16 Chs All four Units [On] [On] 10,000
8861 16 Chs x 2 All eight Units [On] [On] 10,000
If “Too many measurement channels” appears
You have tried to use more channels than the number enabled for use. Either
increase the number of channels to use, or turn unneeded channels [Off].