user manual

3.5.7
3 - 85
(c) Click the Reset button. The following message will appear.
Fig. 3.5-56 Confirming the Operation Method
As described in the message, bring the structure of interest to the image center at a
magnification of about 500×, then click Enter button of Tilt 0 deg.
(d) Stage is moved to tilt angle of 20°. Again bring the structure of interest to the image
center, and click Enter button of Tilt 20 deg. A Calibration Factor is calculated and
indicated. This value represents the deviation of specimen height from the reference
position. Now click the Apply button.
Click the Close button to close the window.