user manual

3.5.7
3 - 89
3) Click the Reset button. The following message indicating steps of operation will
be shown.
Fig. 3.5-62 Verifying the Operating Method
Upon closing the message, the Z-axis will be driven to 30 mm. Carry out focusing
and set magnification to about 500×, then bring the point of interest to the image
center and click Enter button.
4) The Z-axis of stage will be driven to 50 mm. Carry out focusing again, bring the
point of interest to the image center and click Enter.
5) At each click of Enter, the Z-axis will be driven to 35 mm, 20 mm, 10 mm and then
5 mm. Bring the point of interest to the image center each time and then click
Enter.
After clicking Enter at 5 mm, click Apply and then click Close to close the box.